North American Urgent Recall: Neonatal, Pediatric and Flextend Tracheostomy Tubes

North American Urgent Recall: Neonatal, Pediatric and Flextend Tracheostomy Tubes

ID: 99820

(firmenpresse) - OTTAWA, ONTARIO -- (Marketwire) -- 12/23/11 -- The issue:

Health Canada is informing Canadians and Canadian healthcare practitioners that Smiths Medical is conducting a North American voluntary recall of certain lot numbers (see below) of Bivona® Neonatal, Pediatric and Flextend Tracheostomy Tubes. A tracheostomy is an opening in the trachea made to help the patient breathe after surgery, or if there is a blockage in the airway. The kits are used to help keep the opening clear and clean.

According to the company, the recall is due to some customers experiencing difficulty disconnecting accessories from the connectors of the affected tubes. Some customers were also unable to disconnect the accessory or excessive force led to removal of the tube. If the tube is accidentally pulled-out, this can be a life-threatening situation. In some cases, emergency intervention by experts may be required to reinsert the tube to prevent suffocation.

Who is affected:

Canadians who have purchased or used Bivona® Neonatal, Pediatric and Flextend Tracheostomy Tubes.

What consumers should do:

-- Consult your healthcare practitioner if you care for an infant or child that uses this device.

-- Identify all affected unused device in inventory and segregate it to a quarantine location, according to the company.

-- Contact Smiths Medical at 905-477-2000 ext. 135 with any questions regarding the Customer Information Bulletin.

-- View product (), according to the company.

-- Visit the (), complete the "Confirmation Form" and return it to the company by Fax to 905-477-2144 or by email to

What Health Canada is doing:

Canadians can be confident that Health Canada is monitoring the voluntary recall by Smiths Medical. Should additional information be identified, Health Canada will take appropriate action and inform Canadians.

Background:

The tubes are class 2 medical devices and are used on the pediatric population in hospital as well as in the home setting. There has been twenty two complaints concerning the disconnection problem, but none have been reported in Canada.





Details of Affected Devices: ()





For more information:

Consumers and health professionals wanting more information about this advisory from Health Canada can contact the Public Enquiries Line at 613-957-2991, or toll free at 1-866-225-0709.

Media enquiries related to this Advisory should be directed to Health Canada Media Relations at 613-957-2983.

How to report side effects to health products:

To report suspected adverse reaction to these or other health products, please contact the Canada Vigilance Program of Health Canada toll-free at 1-866-234-2345, or complete a and send to us using one of these methods:

Stay connected with Health Canada and receive the latest advisories and product recalls using these social media tools:

Egalement disponible en francais



Contacts:
Media Inquiries:
Health Canada
613-957-2983

Public Inquiries:
613-957-2991
1-866-225-0709

Weitere Infos zu dieser Pressemeldung:

Themen in dieser Pressemitteilung:


Unternehmensinformation / Kurzprofil:
drucken  als PDF  an Freund senden  Minister Kenney Issues Statement Celebrating Christmas Governor General and Commander-in-Chief of Canada Visited Canadian Troops and Civilians Deployed Abroad for the Holidays
Bereitgestellt von Benutzer: MARKETWIRE
Datum: 23.12.2011 - 18:22 Uhr
Sprache: Deutsch
News-ID 99820
Anzahl Zeichen: 3738

contact information:
Town:

OTTAWA, ONTARIO



Kategorie:

Government & Administration



Diese Pressemitteilung wurde bisher 633 mal aufgerufen.


Die Pressemitteilung mit dem Titel:
"North American Urgent Recall: Neonatal, Pediatric and Flextend Tracheostomy Tubes"
steht unter der journalistisch-redaktionellen Verantwortung von

Health Canada (Nachricht senden)

Beachten Sie bitte die weiteren Informationen zum Haftungsauschluß (gemäß TMG - TeleMedianGesetz) und dem Datenschutz (gemäß der DSGVO).

Media Advisory: Health Canada ...

OTTAWA, ONTARIO -- (Marketwired) -- 12/16/13 -- Parliamentary Secretary Eve Adams, on behalf of Rona Ambrose, Minister of Health, along with a toy safety expert from Health Canada`s Consumer Product Safety Division, will hold a demonstration on sma ...

Use a Digital Food Thermometer for Safe Cooking ...

OTTAWA, ONTARIO -- (Marketwired) -- 12/10/13 -- The only reliable way to make sure that meat, poultry, fish and seafood dishes reach safe internal cooking temperatures before serving is to use a digital food thermometer.While we often look for other ...

Alle Meldungen von Health Canada



 

Werbung



Facebook

Sponsoren

foodir.org The food directory für Deutschland
News zu Snacks finden Sie auf Snackeo.
Informationen für Feinsnacker finden Sie hier.

Firmenverzeichniss

Firmen die firmenpresse für ihre Pressearbeit erfolgreich nutzen
1 2 3 4 5 6 7 8 9 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z