Validation of the Memmert hot-air steriliser

Validation of the Memmert hot-air steriliser

ID: 26590

Process validation for a Memmert hot-air sterilisers is no problem at all. As an example, the Schwabach manufacturer of temperature control appliances commissioned the independent and accredited test laboratory biomedis to validate two appliances. The test setup was based on the international standard ISO 20857 for the validation of sterilisation processes by means of hot air, published in August 2010.

 Memmert hot-air steriliser SNE 200 Memmert hot-air steriliser SNE 200

(firmenpresse) - Schwabach, Updated November 2010

In order to protect patients, the requirements made on the quality of medical products are extremely strict, and quite rightly so. For this reason, it was important for Memmert to have the absolute reliability of its hot-air sterilisers verified by an independent organisation. The test setup was based on the guidelines of the DGKH (German Association for Hospital Hygiene) and on the specifications of the standard ISO 20857 for the sterilisation of medical products. The test was performed once without a load, once with mixed load and twice at full load, with some metallic instruments partially wrapped in plastic or aluminium foil as well as glass containers. In the SNE 200 steriliser (with a chamber volume of 32 litres), a total of 16 measurement sensors monitored the constant temperature. Due to the larger volume of 108 litres in the SFP 500 steriliser, 27 sensors were distributed in the chamber and attached directly on the chamber load. The results were identical for both appliances. During each measurement, the minimum temperature of 180 °C was reached and maintained for at least 30 minutes at each measuring point.

Setpoint-dependent sterilisation time
Apart from optimum temperature distribution and temperature stability in the working chamber, a third, crucial requirement is made of hot-air sterilisers: It must be possible to adjust the sterilisation time depending on when the setpoint temperature is reached. In the Memmert hot-air steriliser, the SPWT function, the continuation of the programme depending on the setpoint being reached, ensures that sterilisation does not begin before the predefined sterilisation temperature has been reached. In addition to the standard Pt100 temperature sensor, up to three external temperature sensors can be integrated in the Memmert hot-air steriliser programme sequence and thus ensure that the sterilisation time does not start before the set temperature has been reached at all the measurement points.




Weitere Infos zu dieser Pressemeldung:

Themen in dieser Pressemitteilung:


Unternehmensinformation / Kurzprofil:

Already in its third generation, Memmert GmbH develops and manufactures laboratory appliances with more than 150 employees at two locations in Germany. The product range includes devices such as heating oven, incubator, sterilizer, vacuum drying oven, Peltier-cooled incubator, CO2 incubator, humidity chamber, constant climate chamber, climatic test chamber, temperature test chamber, storage chamber, waterbath and oilbath.

The appliances are used for a very wide range of applications, e.g. material testing, component inspection, biological, chemical and food research, quality testing, human and veterinary medicine.



Leseranfragen:

Angelika Henneberg
Memmert GmbH + Co. KG
Äußere Rittersbacher Str. 38
91126 Schwabach, Germany
Telefon +49 (0)9122 925-0
ahenneberg(at)memmert.com
www.memmert.com



drucken  als PDF  an Freund senden  Attensity?s Multi-Channel Voice of the Customer Text Analytics Now Available In German New licensing model for the InLoox product line
Bereitgestellt von Benutzer: memmert
Datum: 25.11.2010 - 00:03 Uhr
Sprache: Deutsch
News-ID 26590
Anzahl Zeichen: 0

contact information:
Contact person: Angelika Henneberg
Town:

Äußere Rittersbacher Str. 38, 91126 Schwabach, Germany


Phone: +49 (0)9122 925-0

Kategorie:

Manufacturing & Production


Typ of Press Release: Produktinformation
type of sending: Veröffentlichung
Date of sending: 25.11.2010

Diese Pressemitteilung wurde bisher 384 mal aufgerufen.


Die Pressemitteilung mit dem Titel:
"Validation of the Memmert hot-air steriliser"
steht unter der journalistisch-redaktionellen Verantwortung von

Memmert GmbH (Nachricht senden)

Beachten Sie bitte die weiteren Informationen zum Haftungsauschluß (gemäß TMG - TeleMedianGesetz) und dem Datenschutz (gemäß der DSGVO).

Stability tests in the climate chamber ICH ...

Schwabach, May 2011 The new Memmert climate chamber ICH has been specifically designed for stability tests in accordance with GMP and ICH guidelines. In order to uniformly heat or cool the chamber, Memmert deploys an air jacket temperature con ...

NEW! Memmert constant climate chamber HPP 108 with light ...

Schwabach, January 2011 The Memmert constant climate chamber literally is the brightest star of the Schwabach-based company’s product portfolio. In response to requests from many customers, from February 2011 on, Memmert is offering a light modu ...